Flight Dental Systems – Autoclave Error Code Guide
Contact your authorized Flight Dealer for technical support.
Error | Phase | Description | Probable cause | Check | Solution |
E01 | The steam generator overheated. | Temperature sensor disconnected. | Check temperature sensor harness connection on PC board. | It may be necessary to wait up to 10 minutes for temperature to dissipate, then initiate a new cycle.
Secure loose / damaged connection. |
|
Temperature sensor malfunctioning. | Perform: Temperature Sensor Testing | Follow instructions in test procedures. | |||
E02 | The temperature inside the chamber exceeds 137°C. | Temperature sensor disconnected. | Check temperature sensor harness connection on PC board. | It may be necessary to wait up to 10 minutes for temperature to dissipate, then initiate a new cycle.
Secure loose / damaged connection. |
|
Temperature sensor malfunctioning. | Perform: Temperature Sensor Testing | Follow instructions in test procedures. | |||
E03 | The Chamber wall overheated. | Temperature sensor disconnected. | Check temperature sensor harness connection on PC board. | It may be necessary to wait up to 10 minutes for temperature to dissipate, then initiate a new cycle.
Secure loose / damaged connection. |
|
Temperature sensor malfunctioning. | Perform: Temperature Sensor Testing | Follow instructions in test procedures. | |||
E05 | Fail to release the pressure.
Or it is over-pressure. |
Valve (V1) malfunctioning | Inspect two wires connecting the valve and PC board for damage and/or loose connections. | Check for debris in the valve and related tubing / fittings.
Clean / Repair / replace malfunctioning valve. |
|
Sterilizer overloaded. | Check the size of the load. | Advise operator to reduce load size. | |||
Valve (V4) malfunctioning
|
Check if there is a lot of water inside chamber | Check for debris in the valve and related tubing / fittings.
Clean / Repair / replace malfunctioning valve. |
|||
E06 | Door switch disconnected | Check switch harness connection on PC board. | Secure loose / damaged connection. | ||
Check harness connection on the switch. | Secure loose / damaged connection. | ||||
E11 | Steam Generator fails to warm up | Steam generator thermal protector malfunctioning | Check / Inspect the thermal protector | Repair / Replace malfunctioning thermal protector. | |
Steam generator heater malfunctioning | Check / Inspect the wires connecting the heater and PC board for damage and / or loose connections | Repair / Replace malfunctioning heater. | |||
E12 | Chamber fails to warm up | Chamber heater thermal protector malfunctioning | Check / Inspect the thermal protector | Repair / Replace malfunctioning thermal protector. | |
Chamber heater malfunctioning | Check / Inspect the wires connecting the heater and PC board for damage and / or loose connections | Repair / Replace malfunctioning heater. | |||
E04
|
Temperature lower than requested value | ||||
|
Temp<133ºC,Pressure>230Kpa @134 ºC Program
Temp<120ºC,Pressure >130Kpa @121 ºC program: |
Calibration | Check the values of temperature and pressure by and external calibrated device | Run calibration process/Temperature Deviation | |
No steam inside the chamber | Water tank filter blocked | Inspect thewater tank filter. | Clean / replace water tank filter. | ||
Water pump malfunctioning | Inspect the wires connecting the water pump and PC board for damage and/or loose connections. | Repair / replace malfunctioning component. | |||
V3 valve malfunctioning | Inspect two wires connecting the valve and PC board for damage and/or loose connections. | Check for debris in the valve and related tubing / fittings.
Clean / Repair / replace malfunctioning valve. |
|||
Steam generator malfunction | Steam generator not warm . | Replace malfunctioning steam generator | |||
|
Temperature higher than 100 °C and there is water inside the chamber.
|
Valve (V1) malfunctioning
Valve (V4) malfunctioning |
Inspect two wires connecting the valve and PC board for damage and/or loose connections. | Repair / Reconnect / Replace malfunctioning valve. | |
|
During the pressure rising check if there is water and steam as bubbles flowing through the valves. | Check for debris in the valve and related tubing / fittings.
Clean / Repair / replace malfunctioning valve. |
|||
Pressure leaking | Pressure leaking from the hydraulic circuit. | Check for pressure leaks. | Repair / replace malfunctioning component. | ||
E09 | Temperature lower than requested value during holding-time. | Refer to E04 | Refer to E04 | Refer to E04 | |
E21 | Temperature lower than requested value near holiding time | Refer to E04 | Refer to E04 | Refer to E04 | |
E10 | Door locking system malfunctioning. | Electromagnet door locker malfunctioning | Perform: Electromagnet Testing | ||
Inspect two wires connecting the electromagnet lockerand PC board for damage and/or loose connections. | |||||
Switch of Electromagnet door locker malfunctioning | Perform: Switch Testing | ||||
Inspect two wires connecting the switch and PC board for damage and/or loose connections. | |||||
E20 | The cycle is interrupted manually | Shut off the power and restart the machine. |